HANGZHOU CHUNLAI TECHNOLOGY CO., LTD.
TY_ENS
GCOM-3000 Volatile Organic Compounds Online Analyzer

GCOM-3000 Volatile Organic Compounds Online Analyzer

GCOM-3000 Volatile Organic Compounds Online Analyzer

Overview Of GCOM-3000 Volatile Organic Compounds Online Analyzer

GCOM-3000 is a VOCs monitoring product that uses international mainstream chromatographic methods. It can monitor Organic pollutant gases such as methane, the total hydrocarbon, nonmethane total hydrocarbon,low-carbon aldehyde ketone, benzene series (benzene, toluene, xylene, ethylbenzene, isopropylbenzene) and part of halogenated hydrocarbons by using FID detector and high-performance chromatographic separation column. The pneumatic heating box is designed inside the analyzer which can solve the problem of the measured results is low under the certain high-temperature conditions.

Features Of GCOM-3000 Volatile Organic Compounds Online Analyzer


  • After each sample injection, the gas analyzer equipment automatically blows the chromatographic column to speed up the component sieving and analysis;
  • The gas line is inertly processed to avoid the adsorption and residue of samples on the pipe wall;
  • The software in addition to the built-in default analysis method, also allows the user to free configure analysis process and methods;
  • The configuration supports multiple detectors which can be used for complex sample analysis at the same time and can improve the ability of analysis;
  • Support TCP/IP network remote control.

Application Of GCOM-3000 Volatile Organic Compounds Online Analyzer


As one of the best continuous emission monitoring system suppliers in the world, ZETIAN has developed this amazing GCOM-3000 volatile organic compounds online analyzer and it can be applied to emissions of volatile organic compounds monitoring by the end of the manufacturing industry such as crude processing, textile and dyeing, chemical and pharmaceutical, metal smelting, etc.

Specification Of GCOM-3000 Volatile Organic Compounds Online Analyzer


  • Method: GC-FID
  • Range: 0.05ppm~1000ppm(CH4)
  • Distinguishability: 0.001ppm
  • Qualitative Repeatability: RSD7≤1.0%
  • Quantitative Reproducibility: RSD7≤2.0%
  • Range Drift: ≤3.0%(4h)
  • Oven temperature stability: ≤0.5℃
  • Meas.Period: 2mins~30mins (the time will change according to
  • different component separation columns)
  • Dimension: 483mm×266mm×560mm
  • Display: touch LCD screen
  • Comm. Output: 10/100M ethernet port , RS232
  • Comm. Protocol: Modbus-RTU/TCP
  • Power Supply: 220VAC, 50Hz~60Hz, 900W

Product catalog